What buffer should be used to dilute standards, ATP and active PKC in ADI-EKS-420A, PKC Kinase Activity Assay Kit?
Only Kinase Assay Dilution Buffer (80-1486) provided with PKC kinase activity kit (ADI-EKS-420A) should be used to dilute ATP, Standards…
Tue, Jul 18, 2023

How critical is four minutes centrifugation at 13,800 x g following boiling and cooling samples at step 5 in sample handling with Protein A ELISA kit, ADI-900-057?
The recommended centrifugation is 4 min at 13,800 x g. Depending on your centrifuge, you may need to adjust time…
Tue, Jul 18, 2023

Does PACOCF3, BML-ST336 inhibit all isoforms of phospholipase A2 (PLA2)?
PACOCF3 (BML-ST336) inhibits of both Ca2+-dependent cytosolic cPLA2 (IC50=45 µM) and Ca2+-independent phospholipases A2 iPLA2 (IC50=3.8 µM). Unfortunately, we have…
Tue, Jul 18, 2023

What is the difference between Prostaglandin E2 high sensitivity ELISA kit, ADI-930-001 and PGE2 ELISA kit, ADI-900-001?
The PGE2 high sensitivity ELISA kit with catalog number ADI-930-001 was specifically developed to quantify the levels of PGE2 in…
Tue, Jul 18, 2023

Can cells be fixed with 4% paraformaldehyde with the CYTO-ID® Autophagy detection kits (ENZ-51031 and ENZ-KIT175)?
A fixation step may be included when using the CYTO-ID® Autophagy detection kits (ENZ-51031 and ENZ-KIT175), but it needs to…
Tue, Jul 18, 2023

Can CHO host cell protein standards (ENZ-PRT122) be used as internal control in Western Blot or as standards in CHO Host cell protein ELISA kit (ENZ-KIT128)?
CHO host cell protein standards (ENZ-PRT122) can be used in Western blot or in an ELISA. With CHO Host cell…
Tue, Jul 18, 2023

What are the Intra- assay and Inter-assay precision percent coefficient of variation for the concentrations between lots for E. coli host cell protein ELISA kit, ENZ-KIT127 ?
Concerning the E. coli host cell protein ELISA kit (ENZ-KIT127) our intra-assay %CV is
Tue, Jul 18, 2023

Is Autophagy Detection reagent in ENZ-51031 and ENZ-KIT175 compatible with anti-fluorescence quenching agent and fluoromount-G?
The CYTO-ID® Autophagy Detection reagent in ENZ-51031 and ENZ-KIT175 kits is compatible with anti-fluorescence quenching agent and Fluoromount-G.
Tue, Jul 18, 2023

Can the microplate be read following addition of ROS/Superoxide Detection Solution with ROS-ID® Total ROS/Superoxide detection kit, ENZ-51010?
Yes, plates can be read using bottom reading option without removing the ROS/Superoxide Detection Solution and without washing.
Tue, Jul 18, 2023

Is there a requirement for cell lysis with LEADING LIGHT® Wnt Reporter Assay, ENZ-61001
LEADING LIGHT® Wnt Reporter Assay (ENZ-61001) is a cell-based luciferase and does not require lysis and chemiluminescence signals can be…