What are the critical points to pay attention to during preparation of ß-Lactoglobulin solution?
During preparation of ß-Lactoglobulin solution, the following points should be noted: The ß-Lactoglobulin Control is supplied as a lyophilized powder…
Tue, Jul 18, 2023

Is the binding of the protein to the ProteoStat® detection dye reversible?
The interaction between the ProteoStat® dye and the protein aggregates is non-covalent. So, in theory it is a reversible interaction.…
Tue, Jul 18, 2023

Is it possible to distinguish between protein aggregates of different sizes using ProteoStat® Protein Aggregation Assay, ENZ-51023?
The ProteoStat® detection dye does not differentiate aggregate sizes. The dye will bind to a wide range of protein aggregate…
Tue, Jul 18, 2023

Can Cyto-ID® Green autophagy detection reagent from Cyto-ID® Autophagy detection kit, ENZ-51031, be used in combination with ProteoStat® Detection Reagent in ProteoStat® Protein aggregation assay, ENZ-51023?
Since there is no spectral overlap between Cyto-ID® Green Autophagy Detection Reagent and ProteoStat® Detection Reagent, they can be used…
Tue, Jul 18, 2023

What are the applications for ProteoStat® Thermal shift stability assay kit, ENZ-51027?
The ProteoStat® Thermal Shift Stability Assay kit (ENZ-51027) is used for direct monitoring of protein aggregation arising from a thermally-induced…
Tue, Jul 18, 2023

What is the recommended storage condition for ß-Lactoglobulin solution?
The solution of ß-Lactoglobulin can be stored at 4°C for up to a month. For long term storage, it is…
Tue, Jul 18, 2023

Would any protein work with ProteoStat® Thermal Shift Stability Assay kit, ENZ-51027?
When a protein aggregates in response to thermal stress, then the ProteoStat®dye would detect the aggregates. The exact efficiency of detection…
Tue, Jul 18, 2023

Is the ProteoStat® Thermal Shift Stability Assay Kit, ENZ-51027, compatible with the food proteins?
Beta-lactoglobulin is the control provided in the kit. In the food industry, protein aggregation is a desired attribute since aggregates impart…
Tue, Jul 18, 2023

How different is ProteoStat® TS Detection Reagent in ENZ-51027 from ProteoStat® Detection Reagent in ENZ-51023?
ProteoStat®TS Detection Reagent in ProteoStat® Thermal Shift Stability Assay Kit, ENZ-51027, and ProteoStat®Detection Reagentin ProteoStat® Protein Aggregation Assay, ENZ-51023, are…
Tue, Jul 18, 2023

Cells do not exhibit fluorescence after incubation with the detection reagent.
Cell viability is low. Cells should be in log growth phase. Cell samples cannot be kept longer than 6 hours…
Tue, Jul 18, 2023