- What was used as the standard diluent?
Diluents other than the supplied assay buffer may contain interfering substances that can affect the standard curve. - How was the precision of the standard curve?
If the %CV values for the standard curve signal values (e.g. optical density) are consistently above 5%, it may be a good idea to pay particular attention to pipetting technique. If the standard curve signal values were acceptable but the sample precision was not, the problem relates to the sample. Also, see recommendations under “Poor Precision”. - Were the Blank and NSB values subtracted out?
If the net signal values (e.g. optical density) are not used, the signal values will appear higher than those presented in the sample data in the instruction manual. - How were the standard dilutions prepared?
It is important that test tubes of an appropriate size and material are used. Standard dilutions must be properly mixed (e.g. vortexed) while preparing the serial dilutions. It is also crucial that the standard dilutions be prepared and used within the time specified in the product specific instruction manual. Never store unused standard dilutions for a later use.