- Were the wells washed properly?
All wells receive the same treatment during the wash step. If some are washed less than others, this can translate to poor precision. It is important that the plate is washed thoroughly. If plate washing is troublesome, a squirt bottle can be filled with diluted Wash Buffer and all of the wells completely filled from this. The plate contents can be dumped into the sink and shaken to remove excess buffer. This should be repeated for the number of times recommended in the instruction manual. It is important to remember that adding too little Wash Buffer can result in high background, while adding too much is not a problem. The contents of the wells should be aspirated and the plate tapped dry on lint-free paper towels. Plate tapping should consist of a few taps since excessive tapping may lead to plate drying and inconsistent assay results. - Were the wells aspirated sufficiently after the wash steps?
It is very important that as little Wash Buffer as possible remains in the wells after aspiration. Residual buffer can cause dilution of subsequent reagents. After the last wash step, it is a good idea to hit the plate several times over a piece of paper toweling to remove excess buffer. - How were reagents pipetted into wells?
In order to eliminate precision error, customers need to remember to pre-rinse all pipet tips used in the assay. We usually recommend that the customer draw up the liquid into the tip and aspirate it three times prior to addition into the well. Regular pipet calibration and maintenance is also essential to ensure that the tips fit properly and that the correct volumes are dispensed. Be sure reagents are not splashed between wells or outside of the wells during pipeting (especially if using repeater pipets).